Specifications
43
olved Oxygen Sensors Dissolved Oxygen Sensors Dissolved Oxyge
Standard Mounting for
Model D210 Series D.O. Sensors
The Models D210 Series D.O. OxyProbe sensors are intended
for use in a 1
1
/2”sanitary pipe tee. The sensor is secured into
the tee with a standard flange clamp sealed with a flange
gasket. The tip of the sensor is positioned in the middle of
the process flow when it is placed in a standard size tee.
These sensors can also be used in NovAseptic ports. See the
Tip and Hint below for more information.
NovAseptic Sanitary Port
A novel approach to vessel port design,
this NA-connect
®
port from NovAseptic is
sometimes found on vessels in biotech
facilities. This style of port is flush to the
vessel wall to minimize crevices. The port is
installed at a 15° angle to allow for proper
sensor performance. See the cutaway
drawing to the right for typical installation
dimensions.
D215 Installed in NovAseptic Port
The Model D210 or D215 sensor installed in a
NA-connect
®
port offers potential benefits in
biotech applications. The unique design of
the port offers increased drainage around the
port and most effective Clean-In-Place (CIP).
The flush port allows the use of a 19 mm sen-
sor front end, since there is no need for a for-
ward o-ring to seal the sensor into the vessel.
Minimizing or eliminating the cracks or
crevices improves cleanability.
Sensor length = 70 mm
Insertion
Length =
1.4”
36 mm
Cutaway
View
Clamp removed
for clarity
Model 215
D.O. sensor
installed
in a 1
1
/2”
sanitary
pipe tee
1.4"
35.5 mm
0.9"
23 mm
&
&
T I P S H I N T S
T I P S H I N T S
Broadley-James Corporation (U.S. & Canada) Toll-Free: 800-288-2833 E-mail: sales@broadleyjames.com
C
M
Y
CM
MY
CY
CMY
K
43-novaseptic_2013.pdf 1 8/11/14 10:30 AM