BioRobot EZ1 DSP Workstation User Manual 02/2007 3-9
Heating system
The heating system is located under the far end of the
cartridge rack. It holds the heating positions of the reagent
cartridges.
Tray
The tray is located under the racks and collects any drops of
liquid that may fall.
Tray
3.2.2 Pipettor head
The pipettor head is mounted above the worktable and
moves in the Z direction (i.e., up and down) in order to reach
the samples and reagents on the worktable. The worktable
itself moves in the Y direction (i.e., front to back) in order to
present different samples and reagents under the pipettor
head.
The pipettor head contains 6 high-precision syringe pumps
that are connected to tip adapters that can be attached to
filter-tips. The syringe pumps operate simultaneously to allow
aspiration or dispensing of small volumes of liquid
(50–1000 μl) via the filter-tips.
The pipettor head also contains a magnet whose distance
from the attached filter-tips can be varied. This feature allows
the capture of magnetic particles that are present in the
liquid aspirated into the filter-tips.
Located behind the tip adapters is the piercing unit, a row of
6 metal spikes for puncturing the foil that covers the reagent
cartridges.