BioRobot EZ1 DSP Workstation User Manual 02/2007 5-11
5.6.3 Loading the elution tubes, filter-tips, and sample
tubes
1. Remove the tip rack from the worktable.
2. Load the elution tubes into row 1 of the tip rack.
Label the elution tubes before you load them into the
rack. Ensure that the caps of the tubes are removed
before you start the protocol.
3. Place the filter-tips into the tip holders.
Ensure that the tips are positioned correctly in the
holders.
4. Load the tip holders into rows 2 and 3 of the tip rack.
Note: Some protocols require only one row of filter-tips.
Please follow the messages displayed by the LCD.
5. Load the sample tubes into row 4 of the tip rack.
Ensure that the caps of the tubes are removed before you
start the protocol.