Preventive Maintenance
BioRobot EZ1 DSP Workstation User Manual 02/2007 6-3
3. Press β3β to select βClean pierce unitβ.
The piercing unit moves downward.
4. Wipe the piercing unit using a soft tissue moistened with
70% ethanol. The piercing unit is sharp. Double gloves
are recommended.
5. Wipe the piercing unit using a soft tissue moistened with
distilled water.
Piercing unit