Preventive Maintenance
6-4 BioRobot EZ1 DSP Workstation User Manual 02/2007
6. Press “ESC” to return the piercing unit to its original
position.
7. Clean the tray and racks with 70% ethanol and then with
distilled water.
8. Clean the BioRobot EZ1 DSP worktable with 70% ethanol
and then with distilled water.
9. Wipe the other surfaces of the workatable with a diluted
neutral soap solution and then with distilled water.
You can now run another protocol or switch off the
BioRobot EZ1 DSP.
6.2 Daily maintenance procedure
WARNING
Hazardous chemicals and infectious agents [W11]
The waste contains samples and reagents. This waste
may contain toxic or infectious material and must be
disposed of properly. Refer to your local safety
regulations for proper disposal procedures.
After running the last protocol of the day, perform the daily
maintenance procedure:
1. Clean the piercing unit (Section 6.1, page 6-2).
2. Remove used disposable labware and unwanted samples
and reagents, and discard them according to your local
safety regulations.
3. Check that the tray is clean. If necessary, clean it with
70% ethanol, and then with distilled water.
4. Clean the worktable and its racks with 70% ethanol, and
then with distilled water.
5. Wipe the other surfaces of the workstation with diluted
neutral soap solution, and then with water.