Owner manual

18 | P a g e
CONTROL PANEL OVERVIEW
Figure3: Main Control Panel
Power Switch
The main power switch on the control panel controls all power to the workstation and must be in the
on ( I ) position before any of the BACTRONEZ’s systems are powered. The switch will illuminate
when on.
Chamber Gas Light
This light indicates that anaerobic mixed gas is being injected into the workspace chamber.
Main Temperature Control and Green Digital Display
This control is marked SET TEMPERATURE and consists of up and down arrow buttons for
inputting a set point temperature and switching to calibration mode. The green digital display shows
the incubator’s temperature in its default display, accurate to within + 0.1C, as well as the
temperature set point when the arrow keys are being used to make set point adjustments. In
calibration mode, the control and display are used to enter temperature offsets to match the
BACTRONEZ’s displayed temperature to the temperature detected by an independent calibration
(reference) thermometer.