OPERATION (CONTINUED)
SINGLE GAS CONNECTION
Follow these steps to connect a single cylinder of anaerobic mixed gas.
1. Procure a T-shaped brass fitting.
AMG
GAS 1
IN
GAS 2
IN
Figure 14: T-Fitting AMG Figure 15: Single Gas Connections
2. Using attached short lengths of black OD tubing, connect the fitting to the GAS 1 IN and
GAS 2 IN ports on the top left side of the BACTRONEZ. See figure 13 on the previous
page.
3. Connect a barb adaptor on the T-fitting to the AMG gas regulator using the length of black
OD included with the regulator.
Note: Failure to connect the T-fitting to both the GAS 1 and 2 ports will prevent the pass box auto
cycle from purging the pass box.
DUAL GAS CONNECTION
To connect AMG and an inert recharging gas such as, nitrogen
(N
2
), complete the following steps:
1. Connect the AMG regulator’s tubing to the
BACTRONEZ’s GAS 1 IN port.
2. Connect the inert purging gas cylinder regulator line to
the BACTRONEZ’s GAS 2 IN port.
Note: When plumbed in this configuration, the BACTRONEZ will draw twice from the GAS 2 IN
port, and once from the GAS 1 IN port during pass box auto cycles. This is intended to reduce
pass box AMG consumption. The BACTRONEZ will not draw from the GAS2 IN port during
manual gas injections, sleeve assembly purges, or other operations.
Figure 16: Dual Gas Connections