CATALYST OVEN MODEL: CAT180 INSTALLATION AND OPERATION MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.
TABLE OF CONTENTS SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 GRAPHIC SYMBOLS SECTION 3.0 INSTALLATION SECTION 4.0 CONTROL PANEL OVERVIEW SECTION 5.0 PRECAUTIONS SECTION 6.0 OPERATION SECTION 7.0 MAINTENANCE SECTION 8.
1 Section RECEIVING AND INSPECTION Your satisfaction and safety require a complete understanding of this unit, including its proper function and operational characteristics. Read the instructions thoroughly and be sure that all operators are given adequate training before attempting to put the unit in service. Note: This equipment must be used only for its intended application; any alterations or modifications will void your warranty. 1.
2 Section GRAPHIC SYMBOLS Your oven may be provided with a display of graphic symbols to help identify the use and function of the available adjustable components. These symbols, when shown, indicate: 2.1 You should consult your manual for further description or discussion of a control or user item. 2.2 Indicates "AC Power" 2.3 Indicates "Adjustable Temperature" 2.4 Indicates “Manual Control” 2.4 Indicates "Heating" 2.5 Indicates "Over Temperature" 2.6 Indicates “Protective Earth Ground ” 2.
3 Section INSTALLATION Local city, county, or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user. Under normal circumstances this unit is intended for use indoors, at room temperatures between 5 and 40C and no greater than 80% Relative Humidity. The supply voltage should not vary by more than 10% from the data plate rating.
4 Section CONTROL PANEL OVERVIEW Figure 1 4.1 START Switch: The Start switch initiates the catalyst baking cycle. This switch is a push-release momentary action switch. When activated, a timer is set to run for five (5) hours. During this time the oven heats to a minimum of180°C and bakes the catalyst. When the timer has completed the heating cycle, the element is deactivated and the oven is allowed to cool. 4.
5 Section PRECAUTIONS 5.1 Do not place sealed or heat sensitive containers in or on the oven chamber. Keep all flammable materials away from the oven. 5.2 This oven is NOT designed for use in Class I, II, or III locations as defined by the National Electrical Code. 5.3 This oven is not intended, nor can it be used, as a patient connected device.
6 Section OPERATION 6.1 Power Supply: The power supply must be properly grounded (earthed) and correctly sized to match the unit data plate rating. The supply voltage must match the data plate voltage within 10%. When the cord is plugged into the supply, the green CYCLE COMPLETED light should turn ON. 6.2 Insert the catalyst to be baked into the opening and insert it all the way. Do NOT force! If you experience resistance sliding the catalyst into the oven, your catalyst holder may be deformed.
7 Section MAINTENANCE Note: Prior to any maintenance or service on this unit, disconnect service cord from the power supply. 7.1 Cleaning: Clean the oven interior on a regular basis. Clean the inside of the chamber thoroughly using a long swab and a disinfectant appropriate for your application. DO NOT USE chlorine-based bleaches or abrasive cleaners, as they will damage the oven.
8 SECTION PARTS LIST Description 115V 220V Circuit Breaker 1100505 1100505 Cord Set – USA 1800516 1800537 Cord Set – Euro -- 1800541 CTRL Adj.
WIRING DIAGRAM HOT 15A 7850531 CYCLE START to 1750506 OPENS AT 43˚ C TIMER, ERDI-4311, 8520502 -120V TIMER, ERDI-6311, 8520506 -220V 240V WIRING 1 2 3 240V 7030528 RELAY 7030539 SOCKET 4 115V 7030536 RELAY 7030537 SOCKET 9 10 8 7 6 RELAY COIL 5 1750650 SET TO 210˚ C 4650553 RED LAMP 4650554 GREEN LAMP 2350507 (2) 120VAC/125W NEUTRAL 9851011 11