User's Manual
Page 3 of 29
components presents a danger to life!
Only use a perfect power supply cable
A damaged power supply cable may result in a fire or electric shock. When
disconnecting the power supply cable, always do so by holding the plug.
Only use the same type of fuse
T2A/250V
Do not insert any objects into the housing
Objects inserted into the housing may result in damage to the unit or
personal injury.
Do not place any objects on top of the units
Penetrating liquids may result in a fire or electric shock.
Connection
No contact to a patient must occur when handling the cables.
Do not hurt yourself, when moving the display
The display can be tilted backwards and forwards. Please, pay attention
not to hurt yourself, when moving the display. Fingers or small objects may
get stuck at the bottom of the display.
When moving the display up and down (height adjustment), make sure you
do not squeeze your hand or any other object. The minimum distance
between the display edge and the bottom is only 59 mm.
Caution
Incorrect installation may result in extensive damage to property.
Installation should be carried out by trained personnel
When installing your medical electrical system with our products in an
environment with patients, please observe the safety requirements of EN
60601-1 (IEC 60601-1) for "Specifications for the safety of medical
electrical systems" in order to prevent injury to patients and users of your
systems.
Take appropriate measures to particularly ensure that discharge currents
remain below the required limits: Appropriate measures:
- Disconnecting devices for signal input or output unit
- Use of a safety transformer
- Use of additional PE conductor
Only use the signal cables and interface cables specified by the
manufacturer for the installation.
Use power cables with a PE contact. Only insert into sockets with a PE
contact.