Arm Blood Pressure Monitor Instruction Manual Manual version: V1.3 Software version: V1.
Table of Contents 1、Unpacking Inspection..................................................................... 1 2、Packing List.......................................................................................1 3、Safety Precautions.......................................................................... 2 4、Product Composition...................................................................... 3 5、Intended for Use..............................................................................
Thank you for purchasing this Arm Blood Pressure Monitor . The monitor can be used and store measurement results for two users. It uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. This equipment is not suitable for use in locations, where children are likely to be present.
3、Safety Precautions The warnings and illustrations shown in the manual are intended to enable you to use the product safely and correctly, thus preventing harm to you and others, specific meanings of which are shown as follows: Legend, mark and meaning Warning information, refer to the attached document BF-type anti –shock degree for the application part Comply with local regulations Consult the instructions for use.
4、Product Composition This product is composed of the main body and cuff of an Arm Blood Pressure Monitor . 5、Intended for Use The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. 6、Classification of BP (Blood Pressure) ACC/AHA 2017 Hypertension guidelines define categories of BP in adults as below.
7、Product Parts (1) Main body Reserved only (2) Display screen 4
8、Preparation (1) Install the battery 1) Open the battery cover according to the method shown in the figure (as shown in the above figure) 2) Place 4 AAA dry batteries in the battery compartment, and pay attention to the electrode indication of the batteries. Install the battery as indicated in the picture right under this sentence. (2) Battery replacement If the product is not used for a long time (over 3 months), please take out the batteries.
9、Function Setting (1) User mode In standby mode, press the "SET" button to enter the user group selection interface. Then press the "SET" button to switch and select user groups. (2) Year/month/Date/setting In standby mode, press "SET" for about 3 seconds to enter the date setting, and "year" will flash. Press "MEM" to adjust to the desired year, and then press "SET" to confirm the selection. When the "year" is set, it will automatically enter the month setting. At this time, the "month" icon will flash.
(3) Voice on/off Setting (for voice version only, optional function) 1)After the time setting is finished, it will automatically enter the voice setting (the screen will display "SP"). Short press "memory" button to switch ON/OFF sound, "ON" means open voice, "OFF" means close voice, short press "set" button to confirm and select; 2)Voice broadcast content: Attention, please! Please keep silent and relax. Keep the cuff at the heart level.
Thank you. Your blood pressure is: Systolic pressure xxx millimeter of mercury pillar. Diastolic pressure xx millimeter of mercury pillar. Your pulse is xx beats per minute. Measuring result is normal. Thank you. Wish you a good health. ②If your blood pressure is not in normal range: Measuring result is abnormal. Your blood pressure is high/low Could not measure your pulse at this time.
(4) Unit display setting In standby mode, keep pressing "ON/OFF" button for about 10 seconds to enter the unit selection, and press "MEM" to switch mmHg/kPa. Short press "SET" to confirm the selection;The default is mmHg before delivery. kPa mode mmHg mode 10、How to take proper measurements (1) Preparation before measurement --Remove the clothing on the arm. --Always measure in the same arm (generally the left arm). --Remain still and keep quiet during measurement.
--Readings take under the conditions listed below may affect results: Within an hour after dinner, after having wine, coffee, tea, sports; talking, being nervous, being in unsteady mood, bending forward, moving, room temperature dramatically changing during measuring; inside a moving vehicle, repeated and continuous measuring. ( 2 ) Attaching the arm cuff 1)Plug the arm cuff into your monitor by inserting the air plug into the air jack securely until it clicks. 2)Place your hand through the cuff loop.
Note • The bottom edge of the arm cuff should be 0.79-1.18 inches (2 to 3 cm) above the inside elbow. The air tube is on the inside of your arm and aligned with your middle finger.
3)Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely, so it can not move around your arm. 4)To take a measurement on your right arm When you take a measurement on your right arm, the air tube should run along the side of your elbow, along the bottom of your arm. Be careful not to rest your arm on the air tube. Note: Repeated measurement will result in blood congestion in the arm, which will affect the measurement result.
(3)Sitting Correctly To take a measurement, you need to be relaxed and comfortably seated in a room with a comfortable temperature. Place your arm on the table. • Sit in a comfortable chair with your back and arm supported. • Keep your feet flat and your legs uncrossed. • The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.
Note: If you feel uncomfortable during the measurement, press the ON/OFF button immediately to stop measure.When the air pressure is filled to a certain value, the value on the display screen will slowly drop at a certain speed, and the heartbeat symbol will flash. After the measurement is completed, the systolic pressure, diastolic pressure, and pulse measurements will be displayed on the screen. Note : consult your doctor if unexpected readings are obtained.
the measurement data. When the memory is full, the old measurement data will be overfilled with the new one. 2 ) In standby mode, press the MEM button once and the machine will display the average value of the blood pressure measured for the first three times. Press the MEM button again, the first memory will be displayed. Press the MEM button again and the rest memories will be displayed one by one.
11、Contraindications, Precautions, Warnings • DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves. • DO NOT adjust medication based on readings from this blood pressure monitor. Take medication as prescribed by your physician. ONLY a physician is qualified to diagnose and treat high blood pressure. • DO NOT use this monitor on an injured arm or an arm under medical treatment.
• NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician. • To help avoid strangulation, keep the air tube away from infants, toddlers and children. • This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers and children. Stop using this monitor and consult with your physician if you experience skin irritation or discomfort.
• During measurement, make sure that no mobile device or any other electrical device that emits electromagnetic fields is within 12 inches (30 cm) of this monitor. This may result in incorrect operation of the monitor and/or cause an inaccurate reading. • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading. • DO NOT use in a location where there is moisture or a risk of water splashing this monitor. This may damage this monitor.
pressure when you are at home. 2 ) If the cuff is placed in a position above the heart, the measured blood pressure is low, so please measure in the correct posture. Q2. Why is the blood pressure measured at home higher than that measured in hospitals? 1)The anti-hypertensive drugs lose the efficacy. →Please accept the doctor's instruction. 2)The cuff is not twined correctly. If the cuff position is wrong, no arterial signal will be caught, and the measured blood pressure value will be higher.
Q4. When will the measurement be better? Please make measurement in the morning after you get up to urinate or when your body and mood are in stable state. It is better to make measurement at the same time every day. Q5.Why the blood pressure value measured each time is different? 1)When systole each time, the blood pressure will change to some extent. For example, a person with the pulse of 70 beats per minute will have 100,800 blood pressure changes every day.
and grasp and stretch your left and right palms for 15 times repeatedly. Then the extravasated blood can be dissolved, and you can continue the blood pressure measurement. 2 ) Cuff position and twining method. The measured value varies with the cuff size. Particularly, if the cuff is twined round the elbow, you cannot obtain the correct measured value. --Please use the correct cuff twining method for measurement. The arm circumference range of the enclosed cuff is 22~40cm (center of the upper arm).
Er 3 The test result is abnormal. *Trouble shooting Abnormal phenomenon Reason 1. The arm cuff is applied too high. 2. The arm cuff is not applied correctly. 3. The arm cuff is applied too low. 4. You move or talk during a measurement Systolic pressure is high or diastolic is low Blood pressure is too high or too The posture is not the same low during each measurement. The measurement result varies The mental status changes when with the one doctor takes.
14、Maintenance 14.1 Maintenance To protect your monitor from damage, follow the directions below: • Changes or modifications not approved by the manufacturer will void the user warranty. Caution • DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading. 14.2 Storage Keep your monitor and other components in the storage case when not in use. Store your monitor and other components in a clean, safe location. (1)Remove the arm cuff from the monitor.
• To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself. (2)Gently fold the air tube into the arm cuff. Note • Do not bend or crease the air tube excessively. Do not store your monitor and other components: • If your monitor and other components are wet. • In locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach. • In locations exposed to vibrations or shocks. 14.
14.4 Disposal Dispose of your monitor, other components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution. 15、Product Specification Product name Arm Blood Pressure Monitor Display mode Digital LCD display,3 color backlight Measuring Oscillometric Measurement method Measuring part Upper arm Blood Pressure 0mmHg-295mmHg Measuring value (0kPa-39.
Cuff 22-40 cm Service life 5 years Protection Internally powered ME equipment (When using only against electric batteries) shock Class II ME equipment (AC adapter) Temperature 5℃-40℃ condition Working Humidity environment condition 15%-80%RH(non-condensing) Atmospheric 80kPa-105k Pa condition Storage and Strong impact, direct collision, exposure transportation the sunlight or rain should be avoided during environment transportation.
Accuracy Obtains Joy and Health
B Bluetooth version-Bluetooth function test method 1. Use a mobile phone to test the Bluetooth function, and download "nRF Connect" from an Android mobile phone, as shown in the figure below 2.