User's Manual

Thermometer eTE-89B4 User Manual V1.0
17
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
6 V in ISM and
amateur radio bands
between 0,15 MHz and
80 MHz
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
N/A
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
frequency of the transmitter.
Recommended separation distance
P
V
d ]
5.3
[
1
P
V
d
]
12
[
2
P
E
d ]
5.3
[
1
80 MHz to 800 MHz
P
E
d ]
7
[
1
800 MHz to 2.7 GHz
where p is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from
structures, objects and people.
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553