User's Manual

User Manual for All-in-One Health Monitor
Page XIII/69
Safety instructions for SpO
2
measurements
Continuous use of the SpO
2
probe may result in discomfort or pain, especially for those with microcirculatory
problems. It is recommended that the probe should NOT be applied to the same place for over two hours, change
the measurement site periodically and when necessary.
When the ambient temperature is over 35, please change the measuring site every two hours; when the ambient
temperature is over 37, please do NOT use the SpO
2
sensor, as using in high temperatures can cause burns.
Do NOT place the SpO
2
probe on a finger with edema or fragile tissue.
Do NOT put the SpO
2
probe and pressure cuff on the same limb, otherwise the blood pressure measurement may
affect the SpO
2
measurement.
The device is calibrated to display functional oxygen saturation.
Do NOT allow the sensor cable to twist or bend.
Check the SpO
2
sensor and cable before use. Do NOT use a damaged SpO
2
sensor.
When the temperature of the SpO
2
sensor is abnormal, do not use it further.
Remove nail polish or other cosmetic products from the fingernail.
The fingernail should be of normal length.