User's Manual

User Manual for All-in-One Health Monitor
Page XVI/69
Operation procedure:
Step 1: Connect the infrared temperature probe to the connector on the upper side of device with mark of "TEMP".
When Monitor LCD screen displays " ", it means that the probe is connected successfully.
Step 2: Power on the probe by pressing the measuring key on the probe firstly. The user can take measurement when
the probe screen shows " ".
Step 3: Insert the measuring tip of temperature probe into the earhole and press the power on key. One time beep
means the measurement is finished and the result will be displayed on both probe screen and Monitor screen(as
shown in figure 3.1A).
Note:
If the measured temperature is between 37.5 and 42.9, the infrared temperature probe screen shows"
", and beeps for 4 times; If the temperature is lower than 32, "L" will display on both probe screen and
Monitor screen; If the temperature is higher than 42, "H" will display on both probe screen and Monitor screen.
If the transducer detects hardware failure, the infrared temperature probe screen will show "Err" and will not
enter into measuring mode.