User's Manual

User Manual for All-in-One Health Monitor
Page XXVIII/69
Between the left edge of the breast bone and Rib 5
L(Yellow)/
LA(Black)
F (Green)/
LL(Red)
F (Green)/
LL(Red)
2.4 Pressure Accuracy Verification
Operation procedure:
1. Unscrew the M3x6 screw from the battery compartment on the back of the All-in-One Health Monitor, as shown in
figure 2.9.
2. Take a Cuff connector plug from the battery cover, as shown in figure 2.10. (Note: there are two plugs but you will
only need one.)
3. Air Path Connection: Take a piece of air tube (0.5~1m long, Φ8.0mm/Φ4.0mm diameter). Attach the Cuff connector
with a connector plug on to one end of the air tube. Connect the other end to the 3-way connector. Connect the
other 2 ends of the 3-way connector to a pressure meter such as Mercury Sphygmomanometer as shown in Figure
2.12.
4. Connect the Cuff connector end to the NIBP port on the All-in-One Health Monitor.