Pressure Accuracy Verification is a function to inspect the accuracy of pressure measurement by the NIBP module
inside the device. Technician or equipment manager should do pressure accuracy verification every half year or year in
order to check if the pressure measurement still conforms to the requirement of product performance. If the deviation is
beyond the declared specification, it is permitted to return it to factory for repair or calibration.Before verification,
please connect the monitor to a precise pressure meter such as a mercury pressure meter, which is used as the
reference meter.