7. Only monitor a single person at a time.
Warning: Do not use any other adapters than those provided by Creative.
4.7.2 Classification
1. Protection against electric shock: Class II equipment and internally powered equipment
2. The degree of protection against electric shock: Type BF applied part
3. Define apply part: cuff, SpO
2
probe, temperature probe, ECG lead wires (optional).
4. The degree of protection against harmful ingress of liquid: The equipment is IPX2 with protection against ingress of
liquid
5. Electro-magnetic Compatibility: Group I, Class A