7.3 Cleaning and Disinfection of Accessories
It is recommended to clean and disinfect the accessories with a piece of gauze soaked in 75% Alcohol or 70%
Isopropanol.
Do not use damaged accessories.
Accessories cannot be entirely immersed into water, liquid or cleanser.
Do NOT use radiation, steam or epoxyethane to disinfect accessories.
Wipe off any remaining residue of alcohol or isopropanol after disinfection.
Disinfect the temperature sensitive probe with alcohol.
Wipe the thermometer clean with a mild cloth if it becomes dirty.
Wipe the thermometer clean and return to packaging after use.
7.4 Storage
If the equipment will not be used for a long time period of time, wipe it clean and return it to the packaging. Store in a
dry well ventilated place free from dust and corrosive gases.
Storage environment: Ambient temperature: -20°C~60°C