True RMS Digital Multimeter User Manual
AC True RMS
ACV Bandwidth
Display
Overrange indication
Auto Power Off
Polarity
Measurement Rate
Low Battery Indication
Battery
Fuses
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Safety
The term stands for “Root-Mean-Square,” which
represents the method of calculation of the voltage
or current value. Average responding multimeters
are calibrated to read correctly only on sine waves
and they will read inaccurately on non-sine wave or
distorted signals. True rms meters read accurately
on either type of signal.
50Hz to 20000Hz
50,000 count TFTLCD
“OL is displayed
5-30minutes (approximately) with disable feature
Automatic (no indication for positive); Minus (-)
sign for negative
20 times per second, nominal
is displayed if battery voltage drops below
operating voltage
One 7.2 volt (NEDA 1604) battery
mA, μA ranges; 0.8A/1000V ceramic fast blow A
range; 10A/1000V ceramic fast blow
5ºC to 40ºC (41ºF to 104ºF)
-20ºC to 60ºC
Max 80% up to 87ºF (31ºC) decreasing linearly to
50% at 40ºC (104ºF)
<80% 7000ft. (2000meters) maximum.
This meter is intended for origin of installation use
and protected, against the
users, by double insulation per EN61010-1
and IEC61010-1 2nd Edition (2001) to
Category IV 600V and Category III 1000V; Pollution
Degree 2. The meter also meets UL 61010-1, 2nd
Edition (2004), CAN/CSA C22.2 No. 61010-1 2nd
Edition (2004),and UL 61010B-2-031, 1st Edition (2003)
(-4 F to 140 F)