User's Manual

Input Protection: 300V dc or 300V ac rms.
General Specifications
Clamp jaw opening 1.2" (30mm) approx.
Blue tooth 4.0
Display 3-6/7 digits (6000 counts) backlit LCD
Continuity check Threshold 50; Test current < 0.5mA
Diode test Test current of 0.3mA typical;
Open circuit voltage < 3VDC typical
Low Battery indication ’ is displayed
Over-range indication ‘OL’ display
Measurement rate 3 readings per second, nominal
PEAK Captures peaks >1ms
INRUSH 100MS
Temperature sensor Type K thermocouple
Input Impedance 10M (VDC and VAC)
AC response True rms (AAC and VAC)
Operating Temperature 41
o
F to 104
o
F (5
o
C to 40
o
C)
Storage Temperature -4
o
F to 140
o
F (-20
o
C to 60
o
C)
Operating Humidity Max 80% up to 87
o
F (31
o
C) decreasing linearly to 50% at
104
o
F(40
o
C)
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Battery One (1) 9V Battery
Auto power OFF After approx. 15 minutes
Dimensions & Weight 239 x 80 x 49mm; 305g
Safety For indoor use and in accordance with the requirements for
double insulation to IEC1010-1 (2001): EN61010-2-030
EN61010-2-032 EN61010-2-033 Overvoltage Category III 600V,
Pollution Degree 2.
Function Testing Condition Reading
Diode
Forward DCA is approx.1mA,
open circuit Voltage MAX . 3V
Forward voltage drop of Diode
Continuity
Test current MAX. 1.5mA
Buzzer makes a long sound, While
resistance is less than(50)