JPD-FR302 User Manual Infrared Thermometer Version: 1.0 Prepared in August 2015 Shenzhen Jumper Medical Equipment Co., Ltd.
User Instructions Thank you for choosing the infrared thermometer produced by Jumper Medical Equipment Co., Ltd. For proper use of the product, please read this manual before using the product, and follow this manual when operating the product . Before using the product, please read and follow Safety Precautions. Please keep this manual with the product, so that you can refer to this manual in case of need.
Contents Standard Packing List .................................................................... 1 Safety Precautions ......................................................................... 2 Warning ......................................................................................... 4 Symbol Conventions ...................................................................... 5 Product Introduction ....................................................................... 6 Functions and Features .....
FCC Statement This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1)this device may not cause harmful interference. (2)this device must accept any interference received, including interference that may cause undesired operation. NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
Warning:Changes or modiflcations to this unit not expressly approved by the party responsible for compliance could void the user’s authorty to operate the equipment. ISED Statement - English: This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) This device may not cause interference, and (2) This device must a ccept any interference, including interference that may cause undes ired operation of the device.
Unpacking Check Please open the package carefully before use, check that all accessories are contained. Check whether any component is damaged due to improper transportation, and perform installation and operation following this user manual. If any damage or operation problem exists, please contact the dealer or directly contact Jumper Medical Equipment Co., Ltd.
Safety Precautions Before using the thermometer, please carefully read the following precautions: Caution This product cannot be used to replace a monitoring instrument that continuously monitors the body temperature. The temperature probe is a sensitive component of the infrared thermometer, and keep it carefully. Do not discard the exhausted battery at will; take it to a specified environmental protection place for processing, so as to avoid environmental pollution.
After using the product, clean it and keep it in a dust-free, dry, and ventilated place. When using the product, note the safety precautions below: The product is not waterproof, and do not put it into water or other liquid. For the cleaning and disinfection method, see Cleaning and Disinfection. Do not directly contact the top of the temperature probe, where there is a precise temperature sensor at the top.
Warning Warning During measurement, do not forcibly insert the temperature probe into the auditory meatus, which may hurt the auditory meatus. The thermometer is not a diagnostic instrument, and the measurement result is only used as reference for a doctor. It is risky to perform auto-diagnosis and treatment according to the measurement result; follow the doctor's diagnosis. Do not refit the product without permission of the manufacturer, which may damage the product or cause safety problems.
Do not use the thermometer to make measurement for any part of the body except the ear; the measurement result may be incorrect. Keep out of reach of children. Symbol Conventions Description Symbol BF-type application part Caution! Please see this manual. Indicates general prohibitions, where specific content is described following the symbol.
Caution Indicates that the measurement result may be incorrect or the instrument may be damaged in case of wrong usage. Product Introduction (1) Overview The JPD-FR302 infrared thermometer (referred to as thermometer in the following) is a measurement instrument that uses the infrared receiving principle to measure the temperature in the tympanic membrane of the earhole.
(4) Expected Usage This product intends to be used by a medical department or a family to measure the body temperature. (5) Scope of Application The temperature of a measured object is displayed by measuring thermal radiation of the thermal radiation of the earhole. (6) Contraindications Do not use this instrument for an individual having diseases such as otitis media and ear fester. Functions and Features 1.
Conforming to ergonomic design One-key measurement and automatic temperature measurement No interference to daily life, where measurement may be made when a child is asleep 3. Fast Measurement One-second measurement and no waiting 4. High Accuracy Advanced infrared temperature sensing component with high sensitivity Temperature correction program that ensures precise calculation, where the measurement result truly reflects the body temperature 5.
Schematic Diagram of Appearance Measurement Method 1. Precautions before Use 1) This thermometer is applicable to indoor environments. There should be no strong cross-ventilation between the instrument and a measured object during measurement (for example, a fan, an air-conditioner, and a heater blow against each other). 2) This thermometer is sensitive to the environment temperature; do not hold the thermometer for a long time.
5) Do not have emotional outbursts or take acute exercise before measurement. 6) Make measurement 30 minutes later after taking the thermometer into the measurement environment, if there is a temperature difference between the storage environment and the measurement environment. 7) The measured individual is advised to take a more than 10 minutes rest before measurement, if the individual enters the measurement environment from an environment and there is a temperature difference between the two. 2.
to make the auditory meatus straight and the temperature probe aim at the tympanic membrane direction. 3) When measuring the temperature of the ear, press and release the measurement key. The buzzer beeps once, and the display screen displays the measurement value about 1s later. 3.
1) After the measurement is completed, in the power-off state, long press the measurement key for more than 4s but less than 6s to enter the memory query mode, in which you can view previous measurement records. For detailed operations, see Query of 20 Groups of Memory Data in the following table. 2) After the measurement is completed, use a dry, soft cloth to wipe the thermometer, and keep the thermometer in a dry and ventilated place.
Display and Operation Instructions LCD Display Operation Method and Display Instructions Sound and Backlight In the power-off state, press the measurement key for less than 4s. Then, the thermometer displays the One long buzzing sound and 3s green backlight measurement value and starts up. 1. When 32.0C < T < 37.6C, one long buzzing sound and 3s bright green backlight 2. When 37.6C T After the backlight dies out, 42.
and it is automatically identified as the temperature of the object. When the measured Three short buzzing temperature is higher than sounds and 3s green 100C (212F) backlight When the measured temperature is lower than 0C (32F) Query of 20 Groups of Memory Data LCD Operation Method and Display Display Instructions Three short beeps and 3s green backlight Sound and Backlight In the power-off state, press the measurement key for more than 4s but less than 6s. No sound is emitted.
there are 20 groups of memory data in total. If there is no measurement data, the LCD displays only "- - " and the M symbol, and the No sound is emitted. M symbol blinks. C/F Conversion LCD Operation Steps Sound and Backlight Display In the power-off state, press the measurement key for more than 6s but less than 8s to enter the temperature unit conversion mode. Within 5s after releasing the measurement key, press the No sound is emitted.
When there is a read/write error in the EEPROM data or Three short buzzing correction is not completed, sounds and 3s green "ErC" is displayed. In this backlight case, contact the supplier. When the voltage of the battery is lower than 2.51V±0.15V, a low voltage symbol (not blinking) is displayed after the thermometer is powered on. In this case, replace the battery. 16 No sound is emitted.
Battery Replacing Method 1) When replacing the battery, press and hold the upward sign on the battery cover, take off the battery cover, put two AAA batteries into the battery holder according to electrode signs, and close the battery cover. 2) After the thermometer has been used for a period of time, if the display screen displays the low voltage symbol after the thermometer is powered on, replace the batteries with new ones before normally using the thermometer.
Cleaning and Disinfection Device Cleaning: Recommended detergent: * Medical detergent * Household neutral detergent Cleaning steps: 1) Take out the battery before cleaning. 2) If the instrument body is dirty, use a soft and clean cloth to soak neutral detergent or water, wring the cloth and then wipe the instrument body, and immediately use tissue or dry cloth to wipe the instrument body.
You are advised to clean the thermometer weekly, finish cleaning within 3 minutes each time, and clean the thermometer not more than three times each time. Disinfection: Recommended sanitizer: * Isopropanol solution of 70% concentration * Medicinal alcohol of 75% concentration * Sodium hypochlorite solution of 3% concentration Disinfection steps: 1) Use a clean soft cloth to soak a small amount of disinfectant to wipe the instrument body, and immediately dry it.
Maintenance 1) Each time finishing using the instrument, clean the instrument body and the temperature probe and perform operations following Cleaning and Disinfection. 2) Keep the thermometer in a dry, ventilated, dust-free, and pollution-free place, prevent the thermometer from direct solar radiation, and store and transport the thermometer according to corresponding environment requirements.
Troubleshooting Fault Possible Cause Solution The battery level is Use new batteries of extremely low. the same type or Symptom specifications. Failing to start up The batteries are Check whether the installed according to batteries are installed wrong positive and in the battery holder negative electrodes. according to the electrode signs. The instrument is If the instrument is damaged.
Fault Possible Cause Solution after the than 40C or lower than temperature of 10C instrument is 10C. to 40C. "ErC" is There is a read/write Contact the displayed error in internal storage manufacturer for after the data or temperature processing. instrument is correction is not powered on completed. Symptom powered on The lens of the Use a cotton ball to temperature probe is clean the lens of the dirty. temperature probe, to ensure that the lens is not blocked.
Fault Possible Cause Solution The temperature probe is Contact the damaged. manufacturer for There is direct incidence repairing. of strong light. Do not make Symptom The measurement result is higher than a measurement under normal value strong light.
Technical Specifications Product Name Infrared thermometer Model JPD-FR302 Battery Working Internal power supply Mode Working Voltage DC 3V Battery AAA×2 Specifications Battery Life Alkaline dry battery, about 20000 times Running Mode Short-time continuous running Input Power Less than 60mW Display Mode Segment LCD Measurement Time About 1 second Emissivity 0.95 Temperature Unit C/F can be set. Measurement Temperature: 32.0C to 42.2C (89.6F to Range 107.9F) Precision ±0.2C/±0.
Barometric pressure: 86kPa to 106kPa The maximum permissible error of the accuracy of clinical diagnosis test is specified by using the formula below: Accuracy of Clinical Diagnosis Test 0.3C/0.6F (for 95%) where, T1 and T2 are two temperature reading values of two times of measurement of the tested thermometer, and Tref is a fixed reference temperature.
Appendix I:Manufacturer’s Declaration of the EUT Guidance and manufacturer’s declaration – Electromagnetic emissions – for all equipment and systems The Infrared Thermometer JPD-FR302 is intended for use in the electromagnetic environment specifiled below. The customer or the user of the Infrared Thermometer JPD-FR302 should assure that it is used in such an environment.
Guidance and manufacturer’s declaration – Electromagnetic immunity – for all equipment and systems The Infrared Thermometer JPD-FR302 is intended for use in the electromagnetic environment specifiled below. The customer or the user of the Infrared Thermometer JPD-FR302 should assure that it is used in such an environment.
dips, short interruption s and voltage variations on power supply input lines IEC 61000-4-11 (>95% dip in UT) for 0.
The Infrared Thermometer JPD-FR302 is intended for use in the electromagnetic environment specifiled below. The customer or the user of the Infrared Thermometer JPD-FR302 should assure that it is used in such an environment.
where P is Ihe maximum oulput power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters , as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b lnterference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz,the higher frequency range applies.
NOTE 2 These guidelines may not apply in all siluations. Eleclromagnetic propagation is affected by absorption and reflection from slructures, objecls and people. Field strengths from fixed transmitters,such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
distance between portable and mobile RF communications equipment (transmitters) and The Infrared Thermometer JPD-FR302 as recommended below , according to the maximum output power of the communications equipment. Separation distance according to frequency of Rated transmitter maximum m output 80 MHz to 800 MHz 800 MHz to 2.5 GHz power of transmitter 3.5 7 d= [ E1 ] p d = [ E1 ] p W 0.01 0.12 0.23 0.1 0.38 0.73 1 1.2 2.3 10 3.8 7.
Wellkang Ltd Suite B, 29Harley Street, LONDON, W1G9QR,U.K. Shenzhen Jumper Medical Equipment Co., Ltd 5th Floor Building No.34,Baoyuan lndustrial Zone,Xixiang Street,Baoan District,Shenzhen,Guangdong 518102 .P.R. China Phone: +86-755-26696279 Fax:+86-755-26852025 Mail:info@jumper-medical.