User's Manual

19
You are advised to clean the thermometer weekly, finish
cleaning within 3 minutes each time, and clean the thermometer not
more than three times each time.
Disinfection:
Recommended sanitizer:
* Isopropanol solution of 70% concentration
* Medicinal alcohol of 75% concentration
* Sodium hypochlorite solution of 3% concentration
Disinfection steps:
1) Use a clean soft cloth to soak a small amount of disinfectant to
wipe the instrument body, and immediately dry it.
2) Use a clean cloth to soak a small amount of medicinal alcohol of
75% concentration to wipe and disinfect the periphery of the
temperature probe.
Do not disinfect the instrument by means of high-temperature
steam or ultraviolet irradiation, which may damage the instrument or
accelerate aging of the instrument.
You are advised to disinfect the thermometer before and after
using it each time, finish disinfection with 1 minute each time, and
disinfect the thermometer not more than two times each time.
Perform cleaning and disinfection in the following environment:
temperature: +10C to +40C(50F to 104F); humidity 15% to 85%
RH without condensation; barometric pressure: 86kPa to 106kPa.