User's Manual

27
Guidance and manufacturer’s declaration – Electromagnetic
immunity for all equipment and systems
The Infrared Thermometer JPD-FR302 is intended for use in
the electromagnetic environment specifiled below. The
customer or the user of the Infrared Thermometer JPD-FR302
should assure that it is used in such an environment.
Immunity
test
IEC 60601
Test Leve
Complian
ce level
Electromagnet
ic environment
-guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
6 kV
contact
8 kV air
6 kV
contact
8 kV air
Floors should
be wood,
concrete or
ceramic tile. If
floors are
covered
with synthetic
material, the
relative
humidity
should be at
least 30 %.
Electrostatic
transient /
burst
IEC
61000-4-4
2 kV for
power
supply lines
1 kV for
input /
output lines
N/A
Surge
IEC
61000-4-5
1 kV
differentia
mode
2 kV
common
mode
N/A
Voltage
<5 % UT
N/A