User's Manual

32
distance between portable and mobile RF communications
equipment (transmitters) and The Infrared Thermometer
JPD-FR302 as recommended below according to the
maximum output power of the communications equipment.
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of
transmitter
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d=
p][
1
5.3
E
d =
p][
1
7
E
0.01
0.12
0.23
0.1
0.38
0.73
1
1.2
2.3
10
3.8
7.3
100
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHzthe separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.