3 Understanding Your System
Operator’s Manual 3 - 13
3.5.2 Controls and Calibrators
The controls and calibrators are used to verify accurate operation and calibration of the analyzer.
The controls are suspension of simulated animal blood, specifically manufactured to monitor and
evaluate the analysis precision of the analyzer. The controls are prepared with three levels, namely
low, normal and high. Daily use of all levels verifies the operation of the analyzer and ensures
reliable results are obtained. The calibrators are commercially prepared whole-blood products used
to calibrate some parameters (WBC, RBC, HGB, MCV and PLT etc.) of analyzer to build the
metrological traceability of analysis results. For the use and storage of controls and calibrators,
please refer to the Instruction for Use of each product.
All references related to controls in this manual refer to the “controls” and “calibrators” specifically
formulated for this analyzer by Mindary Animal Medical. You must buy those controls and
calibrators from Mindary Animal Medical or Mindary Animal Medical-authorized distributors.
The following models of controls and calibrators are used with the analyzer:
NOTE:
For the specific test parameters and reference values of parameters, see the reference value sheets
of controls and calibrators.
Solution V-6 SR V-6 Solution Reagent
Probe cleanser V-P V-P Probe cleanser
Table 3-5 Reagents
Applicable Channel Model Name
Table 3-6 Information on Controls and Calibrators
Name Model
Control BC-6D
BR60
BC-RET
Calibrator SC-CAL PLUS