Users Manual Part 1

6 Customizing the Analyzer Software
Operator’s Manual 6 - 5
Date/Time (“Menu > Setup > System Setup > Date/Time”)
You can set up date and time on the “Date/Time Setup” screen.
Lab Info. Setup (“Menu > Setup > System Setup > Lab Info. Setup”)
Users may enter necessary laboratory information on the “Lab Info. Setup” screen.
Flag alarm sensitivity (“Menu > Setup > System Setup > Flag Alarm Sensitivity”)
(administrators)
The analyzer provides the following flags for abnormal blood cell morphology.
Table 6-3 Transmission Mode
Item Description
Transmission Mode When you select LIS as a data
channel, check to enable one or
more transmission functions:
•Auto Retransmit
Auto Communicate
Transmit as Print Bitmap
Data
Communicate QC results as
sample result
Only when “ACK Synchronous
Transmission” is enabled, can
you enable the “Auto
Retransmit” function.
Data channel Two data channels are available:
•LIS
•vetXpert
Operators with user level or
administrator level accounts
cannot change the setting
Scattergram
transmitted as/
Histogram transmitted
as
When you select LIS as a data
channel, you can select the
following options from the pull-
down list:
Not to be transmitted
•Bitmap
•Data
/
Item Description
Date Enter the current date /
Time Enter the current time The analyzer uses the 24-hour clock
system
Date Format Select a date format from the pull-
down list
/
Flag Message Meaning Condition
Immature Gran? Possible presence of
immature granulocytes
Presence of excessive dots in immature
granulocyte sensitive region of the
scattergram
RBC Lyse Resistance? Possible presence of RBC
lyse resistance
Presence of abnormally distributed dots
in the WBC sensitive region of the DIFF
scattergram