Users Manual Part 2

12 - 2 Operator’s Manual
12 Service
12.2.2
12.2.3 Replacing the Parts and Components
CAUTION
Only parts supplied by Mindary Animal Medical can be used for maintenance. For any
question, contact Customer Service Department.
12.3 Reagent Management
CAUTION
In case of power failure or upgrade, take out the latex reagent stored in the latex reagent
compartment in time and store it in a refrigeration environment of 2°C to 8°C.
12.3.1 Viewing Reagent Information
On the “Reagent Setup” screen, you may review the expiration dates, use before dates, open dates,
valid days and remaining volumes“
TapMenu > Setup > Reagent Setup” to enter theReagent Setup” screen.
You may review the expiration dates, use before dates, open dates, valid days and remaining
volumes of the analyzing reagents on theReagent Setup” screen.
12.3.2 Replacing the Reagents
Replace the reagent when the reagent runs out, or when the reagent is insufficient or expired
The whole reagent replacing procedure includes the following steps:
1. Read reagent information by swiping RFID card.
2. Install new reagent.
3. TapReplace” button.
Prime Reagent
(Service >
Maintenance >
Reagent)
DS Diluent when reagents are
contaminated or
expired
Replace the residue
reagent in the pipelines
Prime HGB lyse
Prime DIFF lyse
Prime RET diluent
Prime DIFF dye
Prime RET dye
SR Solution Reagent
Table 12-1 Replacing and priming with reagent