12 Service
Operator’s Manual 12 - 3
BIOLOGICAL RISK
After replacing the reagent container/bag, check the tubing connected to the cap assembly
and make sure it is not bent over.
WARNING
While installing or replacing the fluorescent dye bag, hold the upper corners of the bag or the
part under the bag mouth (where the interior tube is located), in order not to extrude the
reagent out.
Read reagent information by swiping RFID card
Follow below instructions:
1. Enter the “Setup” dialog box.
You can enter the “Setup” dialog box in either of the following ways:
When reagent-related errors occur, alarm messages occur in the lower right corner of the
screen. The “Error Information” dialog box prompts and tap “Remove Error”.
Tap the “Reagent Setup” button or tap “Menu > Setup > Reagent Setup” to enter the
Reagent Setup” screen. Select the reagent that needs to be replaced (multiple reagents
can be selected concurrently), and then tapSetup”.
The “Setup” dialog box pops up, and the top part displays the information of the reagent
to be replaced
2. Read reagent information by swiping RFID card.
DS diluent: tear the RFID card from DS diluent and close the RFID card to the swiping
card area.
FD dye and FR dye: close the label on the reagent bag to the swiping card area directly.
LD, LH, DR and SR reagent: close the RFID card on label of the reagent bottle to the
swiping card area.