12 Service
Operator’s Manual 12 - 5
3. Insert the pickup tube of the cap assembly into the new container, and then turn the cap
clockwise until it is secured.
4. Cap the old container with the cap of the new container and dispose of the container properly.
Tap “Replace” button
After swiping RFID card and installing new reagents, tap Replace” button in the “Setup” dialog
box. The analyzer will automatically prime reagent and replace the old reagent.
12.3.3 Replacing the Waste Container
BIOLOGICAL RISK
Remove the waste container cap and replace the waste container only when the power
indicator is not flickering, in order not to make the waste overflow from the container.
If the waste is discharged using waste container, make sure the pickup tube of the waste
container cap assembly is above, and the tube is smooth and not bent.
Follow below instructions:
1. Get an empty waste container, remove the cap and place it next to the one to be replaced.
2. Remove the supporting board under the cap of the old container.
3. Turn the cap counterclockwise and remove the cap assembly from the old container with
caution.
4. Insert the old cap assembly into the new container as vertically as possible, and secure the cap
by turning it clockwise.
5. Install the supporting board under the new containers cap as shown below.
6. Cap the old container with the cap of the new one, and then dispose of the waste properly.