Users Manual Part 2

12 Service
Operator’s Manual 12 - 9
b. Use a sterilized cotton swab dipped with probe cleanser to wipe the surface of the probe
wipe. Then, use a sterilized cotton swab dipped with clean water to wipe the surface until
no blood residue or other residues are visible.
c. Use a sterilized cotton swab dipped with alcohol to wipe the surface of the blood barrier
bracket until no blood residue or other residues are visible.
3. Close the front cover of the analyzer
Gently lower the cover and close it.
CAUTION
Closing the front cover gently to prevent the vibration from damaging internal parts of the
analyzer.
12.5.2 Cleaning the Analyzer Front Cover
BIOLOGICAL RISK
Mindary Animal Medical does not claim the validity of the listed chemicals in infection
control. For effective control of infection, please consult the Infection Prevention Department
of the hospital or the epidemic professionals.
CAUTION
The user shall perform regular cleaning and sterilization to the cover of the instrument.
Use the specified materials to sterilize the instrument only. For any damage to the
instrument or other accidents caused by using materials other than specified, Mindary
Animal Medical will not provide any warranty.
The cleaning and sterilization may damage the instrument to some extent. It is
recommended to perform sterilization only when necessary according to your laboratory
protocol. Remember to clean the instrument before sterilizing.
Do not use any decontamination or cleaning agents which could cause a HAZARD as a
result of a reaction with parts of the instrument or with material contained in it.
If you accidentally spill hazardous material (for example, reagents or samples) on the
instrument, clean the instrument with specified disinfectant. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
Perform regular cleaning on the covers of the analyzer.
Recommended disinfectant: water, 75% ethanol.
Prohibited disinfectant: 3% hydrogen peroxide.
Item Name
1. sample probe assembly
2. Probe wipe
3. Blood barrier bracket