Users Manual Part 2

12 - 10 Operator’s Manual
12 Service
12.6 Preparing to Pack-up
If the analyzer is not to be used for a long time (over 10 days), you should perform this procedure.
Follow below instructions:
1. TapMenu > Service > Maintenance > Fluidics to enter theFluidics” screen.
2. TapPack-up”, and follow the software instruction to complete the pack-up procedure.
12.7 Screen Calibration
If the touch screen does not correctly respond to the positions you touched, perform the procedure
to calibrate the touch screen.
NOTE:
Do not click with the mouse to calibrate the touch screen.
Follow below instructions:
1. TapMenu > Service > Screen Cal.” to enter the touch screen calibration screen.
2. TapScreen Cal.” in the middle of the screen.
3. Tap the black plus sign at the upper left corner of the screen as instructed by the screen display
to start the calibration.
After the calibration is completed, the software displays “Calibration succeeded.” on the
screen.
12.8 Viewing and Exporting Logs
The “Log” screen records all activities of the analyzer. It contributes significantly to searching for
operation history and troubleshooting the analyzer.
The analyzer can save logs of the recent two years. If number of logs exceeds the upper limit, the
latest log will overwrite the oldest one. You can browse and print logs, but cannot delete them.
Administrators and common users have different authorities:
Table 12-2 Log Types
Item Administrator's level Common user’s level
All Logs View all types of logs View the logs for analyzer startup and
shutdown, user logging in and logging out at
the operator's level.
General
Logs
View all operation-related logs at
both administrator's and operator's
levels.
View the logs for analyzer startup and
shutdown, user logging in and logging out at
the operator's level.
Setup
Adjustment
View all setting adjustment logs at
both administrator's and operator's
levels.
Cannot review