Users Manual Part 2

13 - 6 Operator’s Manual
13 Troubleshooting
0x10205 FR Dye expires.
Replace the reagent
FR Dye expires.
Replace the reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x10400 DS Diluent low
volume. Replace the
reagent
DS Diluent low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x10401 LD Lyse low
volume. Replace the
reagent
LD Lyse low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x10402 LH Lyse low
volume. Replace the
reagent
LH Lyse low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
Error ID Error Message Description Solution