Users Manual Part 2

13 Troubleshooting
Operator’s Manual 13 - 7
0x10403 DR Diluent low
volume. Replace the
reagent
DR Diluent low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x10404 FD Dye low
volume. Replace the
reagent
FD Dye low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x10405 FR Dye low
volume. Replace the
reagent
FR Dye low
volume. Replace the
reagent
Follow below instructions:
1. Touch “Remove Error” button and
register new reagent information into
the reagent setup dialog box
displayed.
2. Touch “Replace” to prime reagent
after replacing the reagent container.
3. If error persists, contact Technical
Support Services.
0x40003 Import "Key." file Import "Key." file Follow below instructions:
1. Tap Menu - “Service” - “Advanced
Toolbox” - “Debug Setup” - “Import
Password”, otherwise, reagents
cannot be replaced when built-in
authorization run out.
0x10100 DIL preheating bath
sensor abnormal
DIL preheating bath
sensor abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x10102 Waste cistern floater
status abnormal
WC2 waste cistern
floater status
abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
Error ID Error Message Description Solution