0x10101 Cistern floater
status abnormal
SCI cistern floater
status abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30105 FS baseline
abnormal
FS baseline
abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
0x30106 FS baseline
abnormal
SS baseline
abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
0x30107 FS baseline
abnormal
FL baseline
abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
0x30200 HGB baseline
abnormal
HGB baseline
abnormal
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
0x00300 DIL syringe action
abnormal
Invalid command to
DIL syringe
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.