Users Manual Part 2

13 - 18 Operator’s Manual
13 Troubleshooting
0x30303 Clog RBC sample
preparation
abnormal
Follow below instructions:
1. RBC sample preparation abnormal.
2. Tap the “Remove Error” button to
remove this error.
3. If the error still exists, contact our
customer service department.
0x20200 Power fan error Power fan blocked Follow below instructions:
1. Check whether the power fan is
stuck.
2. If the error persists, contact our
Customer Service Department.
0x20201 Board fan faulty Board radiator fan is
blocked
Follow below instructions:
1. Check if the fan located on the back
of the analyzer main unit is stuck by
any foreign objects;
2. Tap the “Remove Error” button to
remove the error;
3. If the error still exists after restarting
the instrument, contact our customer
service department.
0x20202 Board fan faulty Board radiator fan is
blocked
Follow below instructions:
1. Check if the fan located on the back
of the analyzer main unit is stuck by
any foreign objects;
2. Tap the “Remove Error” button to
remove the error;
3. If the error still exists after restarting
the instrument, contact our customer
service department.
0x20002 Air pressure
detection board
error
Air pressure
detection board
communication
error
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
Error ID Error Message Description Solution