13 Troubleshooting
Operator’s Manual 13 - 19
0x20003 Air pressure
detection board
error
Air pressure
detection board
calibration
parameter error
Follow below instructions:
1. Tap the “Remove Error” button to
remove the error;
2. Switch off and then switch on the
instrument power;
3. If the error still exists after the
restart, contact our customer service
department.
0x30202 Waste channel
abnormal
HGB waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30203 Waste channel
abnormal
HGB waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30204 Waste channel
abnormal
HGB waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30600 Waste channel
abnormal
DIFF waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30601 Waste channel
abnormal
DIFF waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
0x30602 Waste channel
abnormal
DIFF waste channel
clogged
Follow below instructions:
1. Tap the “Remove Error” button to
remove this error;
2. If the error still exists, contact our
customer service department.
Error ID Error Message Description Solution