Table B-8 Repeatability requirements for blood samples
Parameter Range Whole Blood (CV/Absolute
Deviation d*/SD)
WBC (3.50 ~ 4.50)×10
9
/L 3.0%
4.50×10
9
/L 2.5%
RBC 3.50×10
12
/L 1.5%
HGB (110 ~ 180) g/L 1.0%
MCV (80.0 ~ 100.0) fL 1.0%
PLT 100×10
9
/L 4.0%
Table B-9 Carryover requirements for blood samples
Parameters Carryover
WBC 1.0%
RBC 1.0%
HGB 1.0%
HCT 1.0%