User's Manual

Table Of Contents
Monitoring with the TD60 at the CS Physiological Alarms
TMS60 Operator’s Manual 9 - 11
9.2.5.2 Mortara ECG algorithm
Arrh. event Range or Option Default Step Unit
PVCs High 1 to 100 10 1 minute
Pauses/min 1 to 15 8 1 N/A
Asys. Delay 2 to 10 4 1 seconds
Vtac Rate 100 to 200 130 5 bpm
Vtac PVC 3 to 12 6 1 beat
Extreme Tachy Adult: 100 to 300
Pediatric: 160 to 300
Adult: 160
Pediatric: 180
5bpm
Extreme Brady Adult: 15 to 60
Pediatric: 15 to 80
Adult: 35
Pediatric: 50
5bpm
Multif. PVCs Window 3 to 31 15 1 /min
Tachy High Adult: 100 to 300
Pediatric: 160 to 300
Adult: 120
Pediatric: 160
5bpm
Brady Low Adult: 15 to 60
Pediatric: 15 to 80
Adult: 50
Pediatric: 75
5bpm
Pause Threshold RR 1.2, 1.3, 1.4, 1.5, 1.6,
1.7, 1.8, 1.9, and 2.0
2.0 N/A s