User's Manual

Table Of Contents
QT Monitoring Monitoring with the TD60 at the CS
9 - 18 TMS60 Operator’s Manual
9.4 QT Monitoring
A normal ECG waveform (as shown in the following figure) typically includes a sharp and
well defined QRS complexes with consistent spacing between R waves, and an ECG
baseline that is free of noise and artifact.
A normal ECG waveform (for QT monitoring)
The QT interval in an ECG lead is the time interval from the onset of the earliest deflec-
tion in the QRS complex to the end of the T wave. QT monitoring can assist in the detec-
tion of prolonged QT interval syndrome.
9.4.1 Measurement Limitations
At least one measured V-lead must be available in order for the algorithm to
process QT.
QT/QTc values are calculated with 3-leadwire or 5-leadwire ECG cables.
9.4.2 QT Settings
The following table lists the QT settings in the [QT Analysis] section of the [ECG] tab.
ST Point
J Point
ISO Point
R
T
QT Interval
P
Q
S
Options Description Settings*
QT Analysis Enables or disables QT analysis. On, Off
* The factory default settings are in bold.