User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Configuring the TD60 Maintenance Menu
TMS60 Operator’s Manual 10 - 3
3. Tap to exit the [General] menu.
10.2.3 Configuring the Alarms Menu
Select [Alarms] to configure the alarm tone, reminder tone, reminder interval, minimum
alarm volume, and technical alarm priority.
1. In the [Maintenance] menu, tap [Alarms].
The current setting displays to the right of the option.
2. Select the options described in the following table.
Location Changes the hospital and department
names.
N/A
Notch Filter Configures the ECG Notch filter. This
option is used to filter out AC line noise
from the ECG waveform.
50 Hz, 60 Hz, Off
The default is 50 Hz when the
device is not configured for the
US.
ECG Lead Labeling Changes the ECG lead labeling. AHA, IEC
Calibrate ECG Enables or disables the ECG verification. On, Off
Refer to "Verifying the ECG at the
TD60" on page 14 - 4 for details.
SpO
2
Module Changes the SpO
2
module. Masimo, Nonin
SpO
2
Tone Configures the SpO
2
tone. Mode 1, Mode 2
Options Description Settings*
* The factory default settings are in bold.
NOTE
• Mindray recommends the same SpO
2
tone mode be used for the device
within a monitoring area.
Section & Options Description Settings*
Sounds
Style Allows an authorized user to set the alarm tone
pattern.
ISO, Mode 1, Mode 2
* The factory default settings are in bold.