Configuring the TD60 Maintenance Menu
TMS60 Operator’s Manual 10 - 3
3. Tap to exit the [General] menu.
10.2.3 Configuring the Alarms Menu
Select [Alarms] to configure the alarm tone, reminder tone, reminder interval, minimum
alarm volume, and technical alarm priority.
1. In the [Maintenance] menu, tap [Alarms].
The current setting displays to the right of the option.
2. Select the options described in the following table.
Location Changes the hospital and department
names.
N/A
Notch Filter Configures the ECG Notch filter. This
option is used to filter out AC line noise
from the ECG waveform.
50 Hz, 60 Hz, Off
The default is 50 Hz when the
device is not configured for the
US.
ECG Lead Labeling Changes the ECG lead labeling. AHA, IEC
Calibrate ECG Enables or disables the ECG verification. On, Off
Refer to "Verifying the ECG at the
TD60" on page 14 - 4 for details.
SpO
2
Module Changes the SpO
2
module. Masimo, Nonin
SpO
2
Tone Configures the SpO
2
tone. Mode 1, Mode 2
Options Description Settings*
* The factory default settings are in bold.
NOTE
Mindray recommends the same SpO
2
tone mode be used for the device
within a monitoring area.
Section & Options Description Settings*
Sounds
Style Allows an authorized user to set the alarm tone
pattern.
ISO, Mode 1, Mode 2
* The factory default settings are in bold.