User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Troubleshooting Technical Alarm Messages at the TD60
TMS60 Operator’s Manual 12 - 7
Power Low
Battery
Medium A The battery charge is low. Replace with new
batteries.
Critically
Low
Battery
High A The battery charge is
almost depleted.
Battery
Mainten
ance
Required
Medium A The lithium-ion battery is
aging.
Battery
Error
Medium A The lithium-ion battery
communication is error.
Battery
Type
Error
Medium A The battery contact is not
making adequate
connection.
System Device
Error
High A The selftest of the TD60
main board is error.
The selftest of the
parameter module or
Mindray PAN module are
error.
The parameter module
communication is error
or initialization is error.
Restart the TD60. If
the problem persists,
contact your service
personnel.
Real
Time
Clock
Error
High A The real time clock
initialization is error.
Restorin
g Last
Defaults
Failed
Low C Restoring the last default
configuration is error.
Loading
Defaults
Failed
Low C Loading the default
configuration is error.
Measurement
Alarm
message
Alarm
level
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm level is configurable.
** XX is the ECG lead name.