User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Measurement Specifications Product Specifications
A - 14 TMS60 Operator’s Manual
A.8.2 SpO
2
A.8.2.1 Nonin SpO
2
Module
NOTE
• A functional tester or SpO
2
simulator can be used to determine the pulse
rate accuracy.
SpO
2
Measurement range
0% to 100%
Resolution
1%
Accuracy
70-100%: ±3 digits
0-69%: Not specified
Data update rate
≤2s
*Studies were performed to validate the accuracy of Pulse Oximeter with Nonin SpO
2
sensors by contrast with a
CO-Oximeter. In total, 12 adults (7 male & 5 female) aged from 19 to 35 years old were voluntarily involved in this
study. Five are Caucasian, two are Indian, three are Asian, and two are African-American. All of them conform the
clinical study requirements for age, gender, skin and health.
The following table shows the accuracy (A
rms
) for the SpO
2
sensors in four kinds of SaO
2
ranges. For the Fitting
Curve of the SpO
2
sensors, refer to the "Fitting Curve for Nonin SpO2 Sensors" on page A - 15.
SaO
2
range Measured A
rms
Value (7000A, 7000P, 8000AA, 8000AP)
70% to 100%
1.54
70% to 80%
1.41
80% to 90%
1.97
90% to 100%
1.28
PR
Measurement range
20 bpm to 300 pm
Resolution
1 bpm