A - 14 TMS60 Operator’s Manual
A.8.2 SpO
2
A.8.2.1 Nonin SpO
2
Module
NOTE
A functional tester or SpO
2
simulator can be used to determine the pulse
rate accuracy.
SpO
2
Measurement range
0% to 100%
Resolution
1%
Accuracy
70-100%: ±3 digits
0-69%: Not specified
Data update rate
≤2s
*Studies were performed to validate the accuracy of Pulse Oximeter with Nonin SpO
2
sensors by contrast with a
CO-Oximeter. In total, 12 adults (7 male & 5 female) aged from 19 to 35 years old were voluntarily involved in this
study. Five are Caucasian, two are Indian, three are Asian, and two are African-American. All of them conform the
clinical study requirements for age, gender, skin and health.
The following table shows the accuracy (A
rms
) for the SpO
2
sensors in four kinds of SaO
2
ranges. For the Fitting
Curve of the SpO
2
sensors, refer to the "Fitting Curve for Nonin SpO2 Sensors" on page A - 15.
SaO
2
range Measured A
rms
Value (7000A, 7000P, 8000AA, 8000AP)
70% to 100%
1.54
70% to 80%
1.41
80% to 90%
1.97
90% to 100%
1.28
PR
Measurement range
20 bpm to 300 pm
Resolution
1 bpm