User's Manual

Table Of Contents
Introduction Monitoring ECG
7 - 2 TMS60 Operator’s Manual
7.1 Introduction
The chapter describes the ECG monitoring function of the TD60, including skin prepara-
tion and lead placement, ECG Setup, ECG Waveform, and Pacer.
The TD60 can obtain an ECG value by using a 3/5 lead ECG leadwires in conjunction with
the matching skin electrodes.
For details about CS configuration of the ECG parameters, QT analysis, ST analysis, and
Arrhythmia analysis, refer to Chapter 9 Monitoring with the TD60 at the CS.
7.2 Safety
WARNING
Use manufacturer specified electrodes and lead wires.
Make sure the conductive parts of electrodes and associated connectors
for applied parts, including the neutral electrode, do not contact any
other conductive parts including earth.
Periodically inspect the electrode application site to ensure skin quality.
If the skin quality changes, replace the electrodes or change the applica-
tion site.
Do not touch the patient or any device connected to the patient, includ-
ing the bed and gurney, during defibrillation. Otherwise serious injury
or death could result.
CAUTION
Interference from a non-grounded instrument near the patient and elec-
tro-surgery interference can cause problems with the waveform.
NOTE
After defibrillation, the waveform recovers within 10 seconds applied in
accordance with the manufacturer’s instructions for use.