Preparation for Monitoring ECG Monitoring ECG
7 - 8 TMS60 Operator’s Manual
7.3.4.3 Lead Placement: Pacemaker Patients
The recommended lead placement for monitoring a pacemaker patient is as follows.
A pacemaker patient usually requires a different electrode patch placement configura-
tion than a non-pacemaker patient.
Do not place an ECG electrode directly over the pacemaker generator. Place the elec-
trode patches 3 to 5 inches away from the pacemaker generator area. For example, if the
pacemaker generator is located in the right subclavian area, relocate the Right Arm elec-
trode closer in towards the center of the chest.
3-wire lead placement for a pacemaker patient
(AHA)
3-wire Lead Placement for a Pacemaker Patient
(IEC)
5-wire Lead Placement for a Pacemaker Patient
(AHA)
5-wire Lead Placement for a Pacemaker Patient
(IEC)