User's Manual
Table Of Contents
- 1 Safety
- 2 General Product Description
- 3 Getting Started
- 3.1 Unpacking and Checking
- 3.2 Environmental Requirements
- 3.3 Connecting the ECG Leadwire
- 3.4 Installing the Batteries
- 3.5 Powering On the Unit
- 3.6 Understanding Touch Gestures
- 3.7 Basic Operations
- 3.7.1 Understanding the Screen Display Orientation
- 3.7.2 Browsing the Screen Display
- 3.7.3 Switching the Screen Display Orientation
- 3.7.4 Flipping the Landscape Display
- 3.7.5 Displaying the Quick Keys Area
- 3.7.6 Entering the Main Menu
- 3.7.7 Turning the Display Off
- 3.7.8 Turning the Display On
- 3.7.9 Unlocking the Screen
- 3.7.10 Acknowledging the Nurse Call
- 3.8 Using the Pouch
- 4 User Configurations
- 5 Patient Management
- 6 Alarms
- 7 Monitoring ECG
- 7.1 Introduction
- 7.2 Safety
- 7.3 Preparation for Monitoring ECG
- 7.4 Changing the ECG Settings
- 7.5 Understanding the ECG Display
- 8 Monitoring SpO2 (Optional)
- 9 Monitoring with the TD60 at the CS
- 9.1 Introduction
- 9.2 Physiological Alarms
- 9.3 ECG Monitoring
- 9.4 QT Monitoring
- 9.5 ST Monitoring
- 9.6 Arrhythmia Monitoring
- 9.7 SpO2 Monitoring
- 10 Configuring the TD60
- 10.1 Introduction
- 10.2 Maintenance Menu
- 10.2.1 Entering the Maintenance menu
- 10.2.2 Configuring the General Menu
- 10.2.3 Configuring the Alarms Menu
- 10.2.4 Quick Keys Menu
- 10.2.5 Configuring the Defaults Menu
- 10.2.6 Transferring a Configuration
- 10.2.7 Screen Lock Menu
- 10.2.8 Changing the Passwords
- 10.2.9 Changing the Device Name
- 10.2.10 Demo Mode
- 10.2.11 Service Menu
- 11 Battery
- 12 Troubleshooting
- 13 Cleaning and Disinfecting
- 14 Maintenance
- 15 Accessories
- A Product Specifications
- B EMC
- D Symbols and Abbreviations
- E Anomaly
Preparation for Monitoring ECG Monitoring ECG
7 - 10 TMS60 Operator’s Manual
7.3.6 Checking the Paced Status
It is important to correctly set the patient’s paced status before you start monitoring
ECG.
To check the paced status:
■ On the main screen, tap the HR digital area or ECG waveform area to enter the
[ECG] menu.
OR
1. In the main menu, tap [Patient Info].
2. In the [Pacer] field, check the setting of the paced status.
The current paced status setting displays to the right of [Paced].
3. If the paced status setting is not correct, tap [Paced] and select the correct paced
status.
■ When [Paced] is set to [Ye s ] at the TD60, and the pacer pulse is detected, the
symbol displays in the waveform area of the CS screen, and the pace pulse marks
will display on the ECG waveform both at the TD60 and CS.
■ When [Paced] is set to [No] at the TD60, and the pacer pulse is detected, the
symbol displays in the waveform area of the CS screen.
WARNING
• For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to
[No], the CS could mistake a pace pulse for a QRS and fail to alarm when
the ECG signal is too weak. Do not rely entirely on rate meter alarms
when monitoring patients with pacemakers. Always keep these patients
under close surveillance.
• The pacer pulses may be counted as QRS complexes, hence leading to
wrong HR readings or failure to diagnose certain arrhythmia symptoms.
Be sure to keep a close eye on patient’s with pacemaker devices.
• For non-paced patients, you must set [Paced] to [No].
• False low heart rate indicators or false asystole calls may result with cer-
tain pacemakers because of pacemaker artifact such as electrical over-
shoot of the pacemaker overlapping the true QRS complexes.
• In order to minimize the possibility of interference, place electrodes,
leadwires and TD60s as far away from the pacemaker as possible.