Monitoring ECG Changing the ECG Settings
TMS60 Operator’s Manual 7 - 13
2. Tap to exit the [ECG] menu.
7.4.4 Configuring the Pacer
1. In the [Pacer] section of the [ECG] menu, tap the options described in the follow-
ing table.
2. Tap to exit the [ECG] menu.
Speed Selects the waveform sweep speed. 6.25 mm/s, 12.5 mm/s, 25 mm/s
Options Description Settings*
* The factory default settings are in bold.
Options Description Settings*
Paced Selects the paced status. Unspecified, No, Yes
[Unspecified] is only available for
the first time you set the paced sta-
tus.
Refer to "Checking the Paced Sta-
tus" on page 7 - 10 for details.
Markers Selects the pacer indicator.
Line
A 1 cm line shows above each ECG wave-
form each time the pace pulse is detected.
Dot
A 2 mm dot shows above each ECG wave-
form each time the pace pulse is detected.
Line, Dot, Off
Pacer Reject Selects whether or not to reject the pace
pulses.
On, Off
* The factory default settings are in bold.
NOTE
When [Paced] is set to [Yes], the [Markers] and [Pacer Rejection] options
can be available.