Physiological Alarms Monitoring with the TD60 at the CS
9 - 6 TMS60 Operator’s Manual
9.2.4 Factory Default Arrhythmia Alarm Settings
9.2.4.1 Mindray Algorithm
Parameters
Alarm levels*
Activation
State
Factory default alarm
responses
High Medium Low Message
Record
on
Alarm
Print
on
Alarm
Paging
Switch
Asystole
X**
On Off Off Off
VFib/VTac
X
On Off Off Off
VTac
X
On Off Off Off
Vent. Brady
X
On Off Off Off
Extreme
Tachy
X
On Off Off Off
Extreme
Brady
X
On Off Off Off
PVCs/min X
X
X X On Off Off Off
R on T X
X
X X On Off Off Off
Run PVCs X X
X
X Off Off Off Off
Couplet X X X
X
Off Off Off Off
PVC X X X
X
Off Off Off Off
Vent.
Rhythm
X
X
X X On Off Off Off
Bigeminy X
X
X X On Off Off Off
Trigeminy X
X
X X On Off Off Off
Tachy X
X
X X Off Off Off Off
Brady X
X
X X Off Off Off Off
Pacer Not
Pacing
XX X
X
Off Off Off Off
* X indicates available alarm level, — indicates alarm level not available
** The factory default settings are in bold.