User's Manual


B-1
B EMC and Radio Regulatory
Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2.
NOTE
Using accessories, transducers and cables other than those specified may result in
increased electromagnetic emission or decreased electromagnetic immunity of the
device.
The device or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the device or its components should
be observed to verify normal operation in the configuration in which it will be used.
The device needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the requirements of
CISPR.
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of this
monitor.
Other devices that have RF transmitter or source may affect this device (e.g. cell phones,
PADs, PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Emission tests Compliance Electromagnetic environment -
guidance
Radio frequency (RF)
emissions CISPR 11
Group 1 The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.