User's Manual

B-2
RF emissions CISPR 11 Class B The device is suitable for use in all establishments ,
domestic and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes
Harmonic emissions
IEC61000-3-2
Not apply /
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Not apply
If the system is operated within the electromagnetic environment listed in Table Guidance and
Declaration —Electromagnetic Immunity, the system will remain safe and provide the following
essential performance,
Operating mode
Accuracy
Function
Accessories identification
Data stored
Alarm
Detect for connection
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
Mainspowerqualityshould
bethatofatypical
commercialorhospital
environment.