Guidance and manufacturers declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE – SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in
such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Not applicable
3V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the [EQUIPMENT or
SYSTEM], including cables,
than the recommended separation
distance calculated from the
equation applicable to the
frequency or the transmitter.
Recommended separation
distance
d = 1.2 p
d = 1.2 80 MHz to 800 MHz p
d = 2.3 800 MHz to 2.5 GHz p
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol: