User's Manual

Min I(Smart) User Manual
13
IEC 61000-4-2
± 15 kV air
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
synthetic material, the relative humidity
should be at least 30 %.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % UT; 0,5 cycle g) At
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at
0 % UT; 250/300 cycle
N/A
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Mini
I(Smart I) Electric Oral Anesthesia
Injector requires continued operation
during power mains interruptions, it is
recommended that the Mini I(Smart I)
Electric Oral Anesthesia Injector be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE U
T
is the a. c. mains voltage prior to application of the test level.